AQUANAL®
Spectro Turbidity
|
|
|
|
Introduction
The AQUANAL®-plus Spectro turbidity is designed as a compact, easy to use instrument to determine turbidity fast and accurate. A light emitting diode (LED) is used as a light source with a photodetector positioned to detect light scattered by a sample at 90°C to the incident beam. The LED is characterised by maximum long-term stability and monochromatic light emission with minimum power input. The sample chamber (the most critical part of any photometer) is fully sealed, so no water can penetrate into the electronic components. The size of the sample chamber ensures easy cleaning of the light entry surfaces. The AQUANAL®-plus Spectro turbidity may be used as test equipment with software-based calibration adjustment facilities. Battery operation makes the unit suitable for both mobile use and laboratory applications. The AQUANAL®-plus Spectro turbidity is supplied as a fully functional unit complete with accessories and calibration standards in a handy case.
Back to top
Kit Content
AQUANAL®-plus Spectro turbidity with round vials, calibration solutions, accessories, instruction manual, guarantee sheet, 9 V battery in plastic case.
Back to top
Technical Data
- Non-Ratio
- Optics
-LED, (l=875 nm)
- Battery
-9 V-block battery (Life 600 tests)
- Auto-OFF
-Automatic switch-off approx. 5 minutes after last keypress
- Ambient conditions
-5 - 40°C -30 - 90 % rel. Humidity (non-condensing)
Back to top
Product List
| Product # |
Product Name
|
Description |
Add to Cart
|
| 70034 |
Spectro Turbidity |
determination of turbidity / infrared |
|
| 70036 |
Turbidity standards |
2 secondary standards: 1 and 10 NTU |
|
| 70037 |
Turbidity standards |
2 secondary standards: 100 and 1000 NTU |
|
| 70110 |
Turbidity standards |
4 secondary standards: 1, 10, 100 and 1000 NTU |
|
| 70044 |
Sample chamber lid |
for Spectro Turbidity |
|
|
|
CE
- DIN EN 55 022
- DIN EN 61 000-4-2
- DIN EN 61 000-4-8
- DIN EN 50 082-2
- DIN EN 50 081-1
- DIN V ENV 50 140
- DIN V ENV 50 204
| Measuring range |
0-2 FNU |
2-20 FNU |
20-200 FNU |
200-2000 FNU |
| Resolution |
0.1 FNU |
0.1 FNU |
0.1 FNU |
1 FNU |
| Accuracy |
±0.2 FNU |
±1 FNU |
±1 FNU |
±50 FNU |
Back to top
Ratio and Non-Ratio Instruments
Ratio Turbidimeters look at light scatter utilizing more than just the 90° sensor. The quotient of the relationship of the sensors is the resultant NTU value.
| 1) Nephelometric Turbidimeters Function |
2) Schematic Diagram of a Ratio Turbidimeter |
 |
Back to top
Light Scatter Characteristics
Small particles Formazine particles 1.75 - 20 microns Styrene-diphenylbenzene suspension 0.2 - 0.8 microns
Conclusion Different standards, because of their physical properties (shape size and refractive index) do not interact with nepheolometric instruments in the same way all the time. |
 |